KV1.3 - Reproducible Compound Application

icon pl   Patchliner data and applications:p36 1 reproducRec

Application of 5 μM quinidine leads to about 50 % block of the KV1.3 currents (blue). After washout, the current is fully recovered (grey). The lower graph shows corresponding Imax (+40 mV) in the absence and presence of 5 μM quinidine, for two different cells with eight consecutive application and washout steps. The recording lasted over 40 minutes!

 

Back to Overview

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.