2020 - Reliable identification of cardiac liability in drug discovery using automated patch clamp: Benchmarking best practices and calibration standards for improved proarrhythmic assessment
Patchliner and
SyncroPatch 384i (SyncroPatch 384PE a predecessor model) publication in the Journal of Pharmacological and Toxicological Methods (2020)
Authors:
Brinkwirth N., Takasuna K., Masafumi D., Becker N., Obergrussberger A., Friis S., Furukawa Y., Hasegawa Y., Oka T., Ohtsuki A., Fertig N., Stoelzle-Feix S.
Journal:
Journal of Pharmacological and Toxicological Methods (2020) doi: 10.1016/j.vascn.2020.106884
Introduction:
Screening compounds for activity on the hERG channel using patch clamp is a crucial part of safety testing. Automated patch clamp (APC) is becoming widely accepted as an alternative to manual patch clamp in order to increase throughput whilst maintaining data quality. In order to standardize APC experiments, we have investigated the effects on IC50 values under different conditions using several devices across multiple sites.
Methods:
APC instruments SyncroPatch 384i, SyncroPatch 384PE and Patchliner, were used to record hERG expressed in HEK or CHO cells. Up to 27 CiPA compounds were used to investigate effects of voltage protocol, incubation time, labware and time between compound preparation and experiment on IC50 values.