ClC - EC50 for inward and outward directed chloride flux

Clc Figure 2   EC50 Efflux vs InfluxIcon N1   SURFE2R N1 data and applications:

We determined EC50 and relative Vmax for Cl- in an Cl-/H+ exchange assay for both transport directions and found a slight asymmetry in transporter kinetics.

Read more

Background and Assay Description:
Assays for Chloride influx and efflux have been performed using different chloride concentrations ranging from 1 mM to 300 mM on the same sensor.

The vesicles have been equilibrated in Cl- free solution during sensor preparation. For the Cl- influx assay, Clc is activated by a chloride concentration jump. When the Cl- efflux assay is performed, the sensor is first flushed with Cl- containing solution to pre-load the vesicles with chloride. Then a chloride is removed by rinsing the sensor with a chloride free solution (negative concentration jump).

By normalizing each dataset to the peak current obtained for 300 mM chloride, datasets obtained from multiple sensors can be averaged.

Impact of the data:
We have determined EC50 and relative Vmax for Cl- influx and efflux, showing that Influx has a transport rate three times faster than efflux, but a lower apparent affinity.

Due to the experimental design, the slight asymmetry of transporter kinetics may be a result of small differences in the experimental conditions: In the efflux assay, loading of the vesicles with chloride during the experiment generates a membrane potential which might affect the kinetics of Cl- efflux.

Author: Nanion Technologies, in collaboration with Merritt Maduke, Stanford University School of Medicine
Target: EriC, ClcA, ClC-ecl
Family / Type: Chloride Carrier/Channel Family (ClCF, TCDB: 2.A.49)
Mode of transport: H+/2Cl- exchange
Organism: Escherichia coli
Sample: Proteoliposomes from purified protein, LPR 20
Expression: Recombinant overexpression in E.coli
Platform:   Icon N1   SURFE²R N1


Back

Nanion コーポレートブログ

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.