NaV1.5 - Late Current Analysis using the CiPA Protocol

CiPA PE NaV1 5 Pharmacology late currenticon sp96   SyncroPatch 384/768 PE (a predecessor model of the SyncroPatch 384) data and applications:
Cells were kindly provided by Charles River.

Screenshots of the PatchControl 384 software showing NaV1.5 current traces in response to the CiPA voltage step protocol, measured on the SyncroPatch 384PE (a predecessor model of SyncroPatch 384) using whole cell patch clamp methodology and single-hole chips. The NaV1.5 late current was activated by the application of 60 nM ATX-II. The IC50 value of Ranolazine of the late sodium current current was determined as 40.4 µM.

 

Back

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.