CaV1.2 - Current Voltage Relationship

CaV12 CHO CR SP384 IV

icon sp96   SyncroPatch 384PE (a predecessor model of SyncroPatch 384) data and applications:
Cells kindly provided by Charles River.

CaV1.2 expressed in CHO cells recorded on the SyncroPatch 384PE (a predecessor model of SyncroPatch 384). A The screenshot shows the data acquisition and analysis software used on the SyncroPatch 384PE. The online analysis values are shown for a current-voltage experiment. B The raw traces from an example cell elicited by depolarizing steps from -60 mV to 40 mV in 10 mV increments from a holding potential of -80 mV are shown. C The normalized current-voltage plot for an average of 272 cells. A Boltzmann equation fit revealed a V0.5 of activation of -4.8 mV.

 

 

Back

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.