Acetylcholine Receptor Alpha 7 - "Human a7 nicotinic Acetylcholine Receptor on the Patchliner "

icon pl   Patchliner application note:   logo pdf   (0.2 MB)
Cells were kindly provided by Galantos.

Summary:

The human a7 nicotinic acetylcholine receptor expressing cells (stable) from Galantos Pharma GmbH were tested on the Patchliner. Repeated stimulations with 100 μM nicotine were tested to validate the system. In a run, six out of eight cells could be measured. All six cells showed characteristic a7 currents. Average peak currents for the shown experiment were 0.9 ± 0.2 nA (n = 6) which is consistent with manual patch clamp results reported by Galantos Pharma.

返回总览

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.