KV1.5 - Dose response curve of 4-AP

icon sp96   KV1.5 Data 3845PESyncroPatch 384PE (a predecessor model of SyncroPatch 384) data and applications:
Cells were kindly provided by Charles River.

Screenshots of the PatchControl 384 software showing a dose-response curve of 4-AP on KV1.5 stably transfected cells. Measured on the SyncroPatch 384PE (a predecessor model of SyncroPatch 384) using multi-hole chips (4 holes per well), the success rate of completed exeriments  was 100%. The IC50 value of 160 µM corresponds well to literature (IC50 4-AP: 270 µM; Gutman et al., Pharmacological Reviews 57: 473-508, 2005). 

返回总览

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.